Abstract

Recent results in the field of film dosimetry demonstrated that the Green–Saunders equation, a solution of the logistic equation describing phenomena of kinetics of chemical reactions, is the absolute calibration function for all radiochromic film types. Taking advantage of the new opto-electronics-based radiochromic film reading method, which allows real-time measurements of the spectral response of radiochromic films, we confirm that the film darkening is ruled by the Green–Saunders equation independently both from the reading instrument or the choice of the observable used for the calibration. In order to demonstrate it, we exposed an XR-QA2 Gafchromic film to 90Sr/90Y beta rays up to 1400 mGy. Film spectra are recorded in real-time. The calibration is performed by means of two analytic methods: evaluation of the integral under the curves from 500 nm to 645 nm and evaluation of the intensity at 570, 600 and 643 nm. Experimental data fit to the Green–Saunders equation for both methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.