Abstract

ObjectivesThe present study aimed to assess the between day reliability of isokinetic and isometric peak torque (PT) during trunk measurement on an isokinetic device (IsoMed 2000). DesignTest-retest-protocol on five separate days. ParticipantsFifteen healthy sport students (8 female and 7 male) aged 21 to 26. Main Outcome MeasuresPT was assessed in isometric back extension and flexion as well as right and left rotation. Isokinetic strength was captured at a speed of 60°/s and 150°/s for all tasks. ResultsFor none of the assessed parameters a meaningful variation in PT during test days was observed. Relative reliability (ICC = 0.85–0.96) was excellent for all tasks. Estimates of absolute reliability as Coefficient of Variation (CoV) and Standard Error of Measurement (SEM in Nm/kg lean body mass) remained stable for isometric (6.9% < CoV < 9.4%; 0.15 < SEM < 0.23) and isokinetic mode (60°/s: 3.7% < CoV < 8.6%; 0.08 < SEM < 0.24; 150°/s: 6.9% < CoV < 12.4%; 0.10 < SEM < 0.31). In contrast, reliability between familiarization day and day 1 was lower (6.6% < CoV < 26.2%; 0.10 < SEM < 0.65). ConclusionsTrunk strength measurement in flexion and extension or trunk rotation in either isometric or isokinetic condition is highly reliable. Therefore, it seems possible to elucidate changes which are smaller than 10% due to intervention programs when a preceding familiarization condition was applied.