Abstract

A gas discharge strontium vapor laser has been shown to operate with up to 90% of its light emitted at 6.45 &mu;m. We have investigated the use of this laser as a potential stand-alone, tabletop alternative to the FEL for ablation of soft tissue. This custom-made laser currently delivers up to 2.4 watts of average power at 13 kHz pulse repetition rate (range 5-20 kHz). Despite a poor spatial beam profile the laser has been shown to ablate both water and soft tissue. However, current pulse energies (< 185 &mu;J) are insufficient for single pulse ablation even when focused to the smallest possible spot size (130 &mu;m). Instead, the high pulse repetition rate causes the ablation to occur in a quasi CW manner. The dynamics of ablation studied by pump-probe (Schlieren) imaging and macroscopic white light imaging showed micro-explosions but at a rate well below the pulse repetition frequency. Histological analysis of ablation craters in bovine muscle exhibited significant collateral thermal damage, consistent with the high pulse frequency, thermal superposition and heat diffusion. Efforts to increase the pulse energy in order to achieve the threshold for pulse-to-pulse ablation are ongoing and will be discussed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call