Abstract
We investigated the ability of the GDx-Nerve Fiber Analyzer (NFA) to discriminate between normal and early glaucomatous eyes among Korean individuals by reviewing the medical records of 217 consecutive subjects: 61 early glaucoma patients, 68 ocular hypertensive patients, and 88 normal subjects. GDx parameters were compared using ANOVA. The Receiver Operating Characteristics (ROC) curve for each GDx-NFA variable was used to diagnose each parameter, and Pearson correlation coefficients were calculated to assess the association between GDx-NFA parameters and visual field indices in early glaucoma. The best GDx parameters to discriminate between early glaucomatous and normal subjects were the number, maximum modulation, ellipse modulation and inferior ratio (i.e. area under the ROC curve > 0.8). A value for the Number of equal to or greater than 27 was optimal for detecting early glaucoma, with a sensitivity of 80.3% and specificity of 80.7%. In addition, symmetry was positively correlated with the corrected pattern standard deviation (CPSD) among visual field indices in early glaucoma.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.