Abstract

Focusing a wave through heterogeneous media is an important problem in medical ultrasound imaging. In such aberrating media, in the presence of a small number of point reflectors, iterative time reversal is a well-known method able to focus on the strongest reflector. However, in presence of speckle noise generated by many non-resolved scatterers, iterative time reversal alone does not work. In this paper, we propose the use of the echoes coming from moving particles in a flow, such as red blood cells, to generate a virtual point reflector by iterative time reversal. The construction of the virtual point reflector is performed by a coherent addition of independent realizations of speckle coming from moving particles. After focusing on a virtual point reflector, ultrasound images can be locally corrected inside an isoplanatic patch. An application for the correction of power Doppler images is presented. A theoretical analysis shows that this iterative method allows focusing on the point of maximal insonification of the uncorrected beam.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.