Abstract

In 2015, Zika virus (ZIKV) appeared as an emerging pathogen, generating a global and urgent need for accurate diagnostic devices. During this public health crisis, several nucleic acid testing (NAT)-based Zika assays were submitted to the US Food and Drug Administration (FDA) for Emergency Use Authorization. The FDA's Center for Devices and Radiological Health, in collaboration with the FDA's Center for Biologics Evaluation and Research, responded to this Zika emergency by developing and producing a reference panel (RP) for Zika RNA (Zika FDA-RP) suitable for performance assessment of ZIKV NAT-based invitro diagnostic devices. Reference panels are a fundamental tool for performance assessment of molecular tests. The panel is composed of five vials: two different heat-inactivated ZIKV strains (PRVABC59 and FSS13025) in concentrated stocks and three blinded concentrations prepared from those strains. The Zika FDA-RP was shared with developers who had devices in the final stages of validation. Invitro diagnostic developers tested the Zika FDA-RP using the FDA-provided protocol. Depending on sample type, 85% (12/14) of the NAT assays had analytical sensitivities between 500 and 5000 RNA NAT-detectable units/mL (NDUs/mL). One device showed better performance (100 NDUs/mL), and another one showed lower performance (10,000 to 30,000 NDUs/mL). Vials of the Zika FDA-RP are available on request to developers who have interacted with the FDA through the review process.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call