Abstract

This study presents a simple and reliable method for numerical evaluation of the sensitivity of SPR systems. A Windowed Fourier Transform (WFT) based method is applied to extract shifts of SPR interferogram. By applying the WFT, the frequencies consisting of SPR interferogram are extracted, and phase shifts of relevant frequencies are also obtained. The normalized amplitude of the frequencies larger than 0.35 is regarded as the most prominent frequency. The summation of the phase shifts of the most prominent frequencies is regarded as an index to numerically analyzing phase shifts of SPR sensors. Through this method, the small change in the sensing surface can be monitored. By using a homemade SPR system with the proposed calculation method, it can detect the NaCl solutions ranging from 0.00003% to 0.03% (wt%), indicating its great promise for applications in fields such as medicine, food safety, and biotechnology.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.