Abstract

The measurement of plasma renin activity using LC-MS/MS is an attractive alternative to radioimmunoassay techniques. A published method for renin activity by LC-MS/MS uses equipment that is not available in all LC-MS/MS laboratories. Here, we present an offline modification to allow users of LC-MS/MS to perform this analysis without specialist equipment. Samples were prepared in duplicate and incubated for 6.5 h and 24 h. Solid phase extraction was performed offline using Waters Oasis MAX ion-exchange 96-well plate. A method comparison was performed between this assay and a previously published assay using on-line solid phase extraction. The offline method for plasma rennin activity gave similar results to the on-line method across the concentration range. Analytical performance was also found to be comparable and the use of a 24 h incubation for low samples was also deemed unnecessary. The method described is an acceptable alternative to a previously published assay which does not require the use of highly specialist equipment and will be applicable to the majority of LC-MS/MS users.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.