Abstract

A wide range oxygen measurement system with the relative measurement error of less than 1.5% across 4 orders of magnitude of oxygen dynamic range has been developed. We proposed a theoretical evaluation method of the measurement capability of oxygen probes, which points that the relative measurement error (δ) is equal to the reciprocal of the product of signal-to-noise ratio (SNR) and relative sensitivity (S r). A minimum value point in the curve of δ with oxygen concentration is existing due to the opposite trends of SNR and S r Based on the evaluation results, an oxygen probe group was selected scientifically for wide range oxygen measurement. The efficacy of the developed measurement system, featuring adaptive oxygen probe switching, was verified through testing, confirming its compliance with accuracy requirements for measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.