Abstract

Novel atomic force probe scanning microscopy combined with white light interference is proposed, in which the deflection of the cantilever is obtained by measuring the movement of the white light interference central fringe. The relationship between the deflection of the atomic force probe cantilever and the movement of the central fringe is analyzed, and the nonlinearity error correction method of probe cantilever deflection measurement is given. A contact operation mode which combines the displacement of the vertical scanning system and the deflection of the cantilever is presented. The self-developed AFM head can be calibrated directly by the laser interferometer of the vertical scanning system. It can be used to measure high-precision engineering surface, optical surface and micro/nanometer surface. Many experiments on the system for a gauge block surface and hard disk surface measurement are conducted, and standard gratings are used to verify the feasibility and performance of the developed method and system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.