Abstract

To correct for the influence of source configuration on the measured activity of spherical and cylindrical brachytherapy sources, a geometric correction factor was calculated for the Standard Imaging HDR-1000 well-type ionization chamber. A Fortran program modelled each source as a lattice of point sources. Because of the cylindrical symmetry of the well chamber, it could be uniquely modelled by point detectors along the perimeter of the radial plane of the detection volume. Path lengths were calculated and attenuation factors were applied to each source - detector point combination individually. The total dose rate at each detection point was found through a Sievert summation of the point source contributions. For sources with identical activities, a correction factor of was calculated, equal to the ratio of the dose rate of the cylindrical source to that of the sphere. Experimental verification using a Nuclear Associates 67-809 series cylindrical source and an Amersham spherical source yielded a correction factor of .

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.