Abstract
This paper describes an 8-channel gel-free EEG/electrode-tissue impedance (ETI) acquisition system, consisting of nine active electrodes (AEs) and one back-end (BE) analog signal processor. The AEs amplify the weak EEG signals, while their low output impedance suppresses cable-motion artifacts and 50/60 Hz mains interference. A common-mode feed-forward (CMFF) scheme boosts the CMRR of the AE pairs by 25 dB. The BE post-processes and digitizes the analog outputs of the AEs, it also can configure them via a single-wire pulse width modulation (PWM) protocol. Together, the AEs and BE are capable of recording 8-channel EEG and ETI signals. With EEG recording enabled, ETIs of up to 60 kΩ can be measured, which increases to 550 kΩ when EEG recording is disabled. Each EEG channel has a 1.2 GΩ input impedance (at 20 Hz), 1.75 μVrms (0.5-100 Hz) input-referred noise, 84 dB CMRR and ±250 mV electrode offset rejection capability. The EEG acquisition system was implemented in a standard 0.18 μm CMOS process, and dissipates less than 700 μW from a 1.8 V supply.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.