Abstract
PurposeTo evaluate a novel aqueous-based liquid embolic (Embrace Hydrogel Embolic System, [HES]) that has been developed to embolize hypervascular tumors by filling the tumor vascular bed and solidifying into a hydrogel. HES was evaluated for embolization safety and efficacy relative to microspheres in a preclinical rabbit kidney model. Materials and MethodsA renal embolization model in New Zealand white rabbits was utilized. Twenty-four rabbits underwent unilateral kidney embolization via the main renal artery with either HES or 40-μm microspheres. Twenty-two rabbits survived the procedure and were monitored for 2, 12, 17.5, or 26 weeks before sacrifice. All rabbits underwent a repeat renal angiogram before necropsy. HES was evaluated for nontarget embolization, safety, and embolization effectiveness as measured by recanalization and viability of embolized tissue. ResultsBoth embolization materials were found to be safe, with targeted tissue necrosis and absence of nontarget embolization. Prenecropsy angiograms found vascular recanalization in 0/14 (0%) HES-embolized kidneys and in 3/8 (38%) microsphere-embolized kidneys (P = .036). Viable kidney tissue was observed in 2/14 (14%) kidneys embolized with HES and 5/8 (63%) kidneys embolized with microspheres (P = .052). All kidneys embolized with microspheres that showed vascular recanalization had viable tissue on histological sections. HES was observed in vessels as small as 10 μm in diameter in histological analysis. ConclusionsHES provided deep, durable vascular bed embolization that resulted in less recanalization and, on an average, less viable target tissue compared with 40-μm microspheres. No systemic effects or nontarget tissue embolization was identified.
Highlights
A renal embolization model in New Zealand white rabbits was utilized
Viable kidney tissue was observed in 2/14 (14%) kidneys embolized with Hydrogel Embolic System (HES) and 5/8 (63%) kidneys embolized with microspheres (P 1⁄4 .052)
All kidneys embolized with microspheres that showed vascular recanalization had viable tissue on histological sections
Summary
A renal embolization model in New Zealand white rabbits was utilized. Twenty-four rabbits underwent unilateral kidney embolization via the main renal artery with either HES or 40-μm microspheres. Twenty-two rabbits survived the procedure and were monitored for 2, 12, 17.5, or 26 weeks before sacrifice. All rabbits underwent a repeat renal angiogram before necropsy. HES was evaluated for nontarget embolization, safety, and embolization effectiveness as measured by recanalization and viability of embolized tissue
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.