Abstract

The design and performance of a 95 GHz pulsed W-band EPR/ENDOR spectrometer is described with emphasis on the ENDOR part. Its unique feature is the easy and fast sample exchange at 4.2 K for frozen solution and single crystal samples. In addition, the microwave bridge power output is relatively high (maximum 267 mW), which allows the application of short microwave pulses. This increases the sensitivity in echo experiments because of the broader excitation bandwidth and the possibility of employing short pulse intervals, as long as the dead time does not increase significantly with the power. The spectrometer features two microwave and radiofrequency (0.1–220 MHz, 3 kW pulse power) channels and a 6 T superconducting magnet in a solenoid configuration. The magnet is equipped with cryogenic sweep coils providing a sweep range of ±0.4 and ±0.2 T for a center field of 0–4 and 4–6 T, respectively. The spectrometer performance is demonstrated on Cu(II) centers in single crystals, a zeolite polycrystalline sample, and a protein frozen solution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.