Abstract

We have developed a setup that offers a vacuum compatible, vibration-free experimental space with active thermal stabilization around room temperature, suitable for high sensitivity measurements. In this paper we describe the experimental setup from the thermal viewpoint and the design and implementation of the thermal control. Then we characterize the long-term performances of the control loop by using experimental data collected over 1 month: we show that our control is effective in reducing room temperature variations by a factor 100 over a few weeks, even in the presence of an internal dc heat source of 2 W.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.