Abstract

The GTS ion source, operated at 14.5GHz, provides multiply charged heavy ion beams for the ARIBE facility at GANIL. The facility variety is limited by the efficiency of the extraction especially in the few keV/q domain. In order to improve the ion source, the extraction system was upgraded numerically using IBSimu. The shape of the plasma and the puller electrodes was changed. It causes the increase of the electric field near the plasma meniscus and allows the use of lower extraction voltages at constant total ion beam current. The required beam parameters were taken from the experimental data. The opportunity of effective low energy beam formation (at a few keV/q or several hundreds eV/q beam energy) was studied. The high current and low energy ion beam production will provide new possibilities for ARIBE facility users.