Abstract
The standard Akta Explorer high-performance liquid chromatography (HPLC) system has limitations for the automation of multidimensional protein purification. Here, we describe simple modifications that allow for automated multidimensional purification protocols to extend the possibilities of the Akta three-dimensional purification kit in terms of column number, flexibility of volumes stocked for re-injection of samples, and available choice of buffers. These modifications do not preclude the use of standard one-dimensional purification protocols. Additionally, we demonstrate a technology for encrypted full remote control of the machine over the Internet by cost-effective use of standard asymmetric digital subscriber line (ADSL) that enables direct remote interaction with the machine without preventing local control. A 4-column purification scheme, including equilibration and cleaning in place (CIP) procedures, was implemented on such a system. It significantly increased reproducibility and shortened processing time by 85%, as compared with manual operation, thus allowing for automated protein purification overnight.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.