A simple, isocratic, rapid, and accurate reversed-phase high-performance liquid chromatographic method has been established for quantitative determination of zonisamide. The method is also applicable to determination of related substances in the bulk drug. Chromatographic separation was achieved on a 250 mm × 4.6 mm, 5-μm particle, C18 column; the mobile phase was a 70:30 (v/v) mixture of 0.1% (v/v) aqueous triethylamine, adjusted to pH 2.5 with dilute orthophosphoric acid, and acetonitrile. Chromatographic resolution of zonisamide from its potential impurity, A, was found to be >2. The limits of detection and quantification of zonisamide and impurity A were 0.04 and 0.12 μg mL−1, respectively, for 20 μL injection volume. Recovery of zonisamide ranged from 98.5 to 101.2% and recovery of impurity A from a sample of zonisamide ranged from 97.4 to 102.7%. The method was validated for linearity, accuracy, precision, and robustness.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.