Abstract

An improved method based on tandem solid phase extraction (SPE) cleanup and gas chromatography-high resolution mass spectrometry (GC-HRMS) has been validated for a rapid determination of dibenzo-p-dioxins/furans (PCDD/Fs), dioxin-like polychlorinated biphenyls (PCBs), marker polychlorinated biphenyls (M-PCBs), and polybrominated diphenyl ethers (PBDEs) using a large volume (50mL) of human milk. This method was well validated for the measurement of these analytes in human milk from the general population with low limits of detection (LODs, 0.004-0.12ng/g lipid), satisfactory accuracy (75-120% of recoveries), and precision [less than 10% of relative standard deviations (RSDs)]. To comprehensively evaluate the performance of this method, a good, presently validated and routinely used method based on an automated sample clean-up system (ASCS, based on the commercial acid multilayer silica, basic alumina, and carbon columns) was used in parallel for comparison. Compared with the ASCS method, this method presented comparable specificity. Additionally, this method, in contrast to ASCS method, highly reduced consumption of solvents (40mL versus 500mL), which results in much lower background in the procedural blank, reduced time, and enhanced sample pretreatment throughput. This method was also applied in a pilot study to measure a batch of human milk samples with satisfactory results. Graphical Abstract Characteristics of the application of tandem SPE cleanup for determination of PCDD/Fs, DL-PCBs,M-PCBs and PBDEs in human milk.