Abstract

An enantioseparation high performance liquid chromatographic (HPLC) method was developed and validated to determine d-(+)- and l-(−)-vesamicol in human plasma. The assay involved the use of a solid phase extraction for plasma sample clean up prior to HPLC analysis utilizing a C18 Bond-Elute column. Chromatographic resolution of the vesamicol enantiomers was performed on a vancomycin macrocyclic antibiotic chiral stationary phase (CSP) known as Chirobiotic V with a polar ionic mobile phase (PIM) consisting of methanol:glacial acetic acid:triethylamine (100:0.1:0.05 (v/v/v)) at a flow rate of 1.0 ml/min and UV detection set at 262 nm. All analyses were conducted at ambient temperature. The method was validated over the range of 1–20 μg/ml for each enantiomer concentration ( R 2>0.999). Recoveries for d-(+)- and l-(−)-vesamicol enantiomers were in the ranges of 96–105% at 3–16 μg/ml level. The method proved to be precise (within-run precision ranged from 1.3 to 2.7% and between-run precision ranged from 1.5 to 3.4%) and accurate (within-run accuracies ranged from 0.8 to 3.4% and between-run accuracies ranged from 1.7 to 5.0%). The limit of quantitation (LOQ) and limit of detection (LOD) for each enantiomer in human plasma were 1.0 and 0.5 μg/ml ( S/ N=3), respectively.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call