Abstract
A stability indicating, reversed-phase high-performance liquid chromatographic method was developed and validated for the determination of clopidogrel in pharmaceutical dosage forms. The determination was performed on a semi-micro column, BDS C8 (250×2.1 mm i.d., 5 μm particle size); the mobile phase consisted of a mixture of 0.010 M sodium dihydrogen phosphate (pH 3.0) and acetonitrile (35:65, v/v), pumped at a flow rate 0.30 ml min −1. The UV detector was operated at 235 nm. The retention times for clopidogrel and naproxen, which was used as internal standard, were 3.08 and 6.28 min, respectively. Calibration graphs are linear ( r better than 0.9991, n=6), in concentration range 1.00–3.00 μg ml −1 for clopidogrel. The intra- and inter-day RSD values were less than 1.96%, while the relative percentage error E r was less than 2.0% ( n=5). Detection and quantitation limits were 0.12 and 0.39 μg ml −1, respectively. The method was applied in the quality control of commercial tablets and content uniformity test and proved to be suitable for rapid and reliable quality control.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.