A simple and sensitive liquid chromatography with ultraviolet detection (LC–UV) method was developed for the determination of three impurities with a content over 0.1% (w/w) in technical triadimefon. A Gemini C18 column (5 μm, 250 mm × 4.6 mm i.d.) was used for the chromatographic separations. The samples were separated by gradient elution with water (solvent A) and methanol (solvent B) using the following conditions: 70% A isocratic for 12 min, linear to 0% A within 8 min, and isocratic for 10 min at 0% A with a flow rate of 1.0 mL min−1. Chromatograms were recorded at an absorption wavelength of 280 nm. The chromatographic resolutions between triadimefon and its potential impurities A, B, and C were greater than 3. The developed LC method was validated with respect to linearity, accuracy, precision, and robustness. This method was successfully applied to analyze the impurities in commercial technical triadimefon. In addition, the structures of the three impurities were identified to be (A) 4-chlorophenol, (B) 1-(2,4-dichlorophenoxy)-3,3-dimethyl-1-(1H-1,2,4-triazol-1-yl)-2-butanone, and (C) 1,1-bis(4-chlorophenoxy)-3,3-dimethyl-2-butanone.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.