A rapid isocratic chiral HPLC method has been developed for the separation of R-galantamine from S-galantamine. Good resolution viz. Rs > 3 between R- and S- forms of galantamine was achieved with chiralpak AD-H (250 × 4.6 mm) column using n-hexane, 20% propionic acid in isopropanol and diethyl amine in the ratio of 80:20:0.2 (v/v) as mobile phase at ambient temperature. Flow rate was kept as 0.8 mL min−1 and the elution was monitored at 289 nm. This method was further used to determine the amount of R-galantamine in pure and pharmaceutical formulations of S-galantamine. This method is capable to detect and quantitate R-galantamine to the levels of 0.21 and 0.84 μg mL−1, respectively. The method was validated as per International Conference on Harmonization (ICH) guidelines in terms of limit of detection (LOD), limit of quantification (LOQ), linearity, precision, accuracy, specificity, robustness and ruggedness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.