Abstract

A gradient reversed-phase liquid chromatographic assay was developed for the quantitative determination of the non-steroidal anti-inflammatory drug valdecoxib. The developed method was also applicable to the determination of related substances in the bulk drug. Forced degradation studies were performed on bulk valdecoxib using acid (2.0 N hydrochloric acid), base (2.0 N sodium hydroxide), oxidation (6.0% v/v hydrogen peroxide), water hydrolysis, heat (60 °C) and photolysis. Mild degradation was observed using alkaline conditions and considerable degradation observed during oxidative stress. Chromatographic separation of process-related impurities and degradation products was achieved using a 5 micron Zorbax SB-CN LC column. The mobile phase consisted of aqueous potassium dihydrogen phosphate (pH 3.0) and acetonitrile. Stressed samples were assayed using the developed LC method and determination of the mass balance accounted for 99.5%, thus indicating the suitability of this stability-indicating method. Linearity, accuracy, precision and robustness have also been evaluated.