Abstract
We describe the design, characterization, and implementation of a vacuum ultraviolet (VUV) monochromator for use in filtering stray and scattered light from the principal monochromator output of the Stainless Steel Seya VUV synchrotron beam line at the Synchrotron Radiation Center, University of Wisconsin—Madison. We demonstrate a reduction of three orders of magnitude of stray and scattered light over the wavelength range 1400–2000Å with minimal loss of light intensity, allowing for over six orders of magnitude of dynamic range in light detection. We suggest that a similar filtering scheme can be utilized in any variety of spectroscopic applications where a large dynamic range and low amount of background signal are of import, such as in transmittance experiments with very high optical density.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.