Abstract

In France, chemotherapy preparation units of hospital pharmacy compound cytotoxic infusion bags adapted to each patient. The narrow therapeutic index of these preparations led us to implement qualitative and quantitative control for patients’ safety. To this aim, we calibrated an equipment combining UV–vis spectrometry and Raman spectroscopy (QC Prep+®) and monitored 14 different molecule–solvent combinations over a 18 months period. This rapid and specific method allowed the qualitative and quantitative analysis of 1mL sample tests in less than 2min. On 5742 anticancer preparations, we obtained accepted results with more than 99.4% solvent identification, 99.6% drug identification and only 1.52% of preparations not matching quantitative specifications (±15% of theoretical concentration). This quantitative control enabled us to pinpoint some critical points of production for two of the most common preparations. We thus updated the procedures of reconstitution and preparation, increasing the quality of final product. UV-Raman spectrometry is thus an effective tool to control chemotherapy infusions and to improve good practices of preparation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.