Abstract

Several glycoconjugate-based vaccines against bacterial infections have been developed and licensed for human use. Polysaccharide (PS) analysis and characterization is therefore critical to profile the composition of polysaccharide-based vaccines. For PS content quantification, the majority of Ultra High Performance Liquid Chromatography (UHPLC) methods rely on the detection of selective monosaccharides constituting the PS repeating unit, therefore requiring chemical cleavage and tailored development: only a few methods directly quantify the intact PS. The introduction of charged aerosol detector (CAD) technology has improved the response of polysaccharide analytes, offering greater sensitivity than other detector sources (e.g., ELSD). Herein, we report the development of a universal UHPLC-CAD method (UniQS) for the quantification and quality evaluation of polysaccharide antigens (e.g., Streptococcus Pneumoniae, Neisseria meningitidis and Staphylococcus aureus). This work laid the foundation for a universal UHPLC-CAD format that could play an important role in future vaccine research and development helping to reduce time, efforts, and costs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call