Abstract
Bile acids malabsorption (BAM) is encountered in numerous gastrointestinal pathologies and is a good example of a treatable cause of watery diarrhea after ileal resection. The gold standard for diagnosing BAM is the selenium homocholic acid taurine test (SeHCAT), an expensive and complex analysis. An alternative method is the quantification of 7α-hydroxy-4-cholesten-3-one (C4). Here, we present a simple, ultra high-performance liquid chromatography–tandem mass spectrometry method to measure C4 in human serum. To avoid time consuming sample preparation (e.g., derivatization, solid phase extraction), we used absorption chemistry-based extraction plates. This method demonstrates a lower limit of quantification of 5ng/mL and is linear over a concentration range from 5 to 300ng/mL (R2=0.9977). Inaccuracy and imprecision were less than 15%. The validated method is currently used for routine measurement of C4 from serum in patients to confirm BAM diagnosis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.