Abstract
A portable micro energy dispersive X-ray fluorescence spectrometer with Laser Displacement Sensor (LDS) is reported in this paper. Due to irregular or curved surface of archaeological object, the distances between the irradiation spot of sample and exit of polycapillary X-ray optics can vary. In this paper, we employed a LDS in our portable micro-X-ray fluorescence spectrometer to control the distances between the irradiation spot on the sample and the exit of polycapillary X-ray optics in order to decrease the errors caused by the irregular or curved surfaces of samples. In order to test this spectrometer, elemental mapping analyses were carried on with and without LDS on an irregular colored glaze of a piece of curved ancient porcelain, the height difference of two parts of ancient porcelain being nearly 5 mm. From the results, it can be concluded that this spectrometer has potential application prospects in measuring irregular or curved surface of samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.