Abstract

We have developed and tested a thermal-neutron, two-dimensional position-sensitive detector for use in neutron scattering from biological samples. The detector is based on a multi-wire proportional concept with a gas filling of 6 atm 3He, 4 atm A, and 0.5 atm CO 2. The position readout uses a resistive wire with charge division to determine position. We have measured a resolution of less than 3 mm (fwhm) with a detection efficiency of 70% at 2.8 Å neutrons. Considerations leading to an optimal configuration are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.