Abstract

Methylglyoxal (MGO), a dicarbonyl compound in living organism, food and environment, has been associated with disease diagnosis and human health. The current electrochemical detection methods rely on the use of advanced materials. In this work, a non-advanced materials “two-step” assay including electrode electro-activation and MGO detection was developed. In the section of electro-activation, an activation method of GCE for MGO detection was established; and the composition changes on GCE surface caused by electro-activation, including functional groups and surface defects, have been carefully studied. The effect of carbonyl and surface defects induced by electro-activation on MGO detection was discussed. In section of MGO detection, the raise of background current caused by electro-activation was minimized by background subtraction; and the effect of interferences can be weakened by adjusting pH. The MGO signal on proposed activated GCE improved 20-fold than bare GCE. The recoveries were 72.38–109.16% in honey and beer, and RSDs were 0.24–9.63% without significant difference with HPLC method and comparable with advanced material modified sensors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.