Abstract

We describe the design, construction and performance of a double stage 3He– 4He refrigerator, built to cool down a multiband bolometric photometer at the MITO telescope. The fridge was optimized to work without external pumps, with the main cryostat providing a 4.2 K thermostat at sea level and a 4.0 K one at high mountain pressure conditions. The measured ultimate temperature of the fridge is 290 mK, with a hold time of 81 h. The external heat input on the cold flange is ∼35 μW, with the main bath at 4.0 K. The recycle time is 8 h with a heat input on the thermostat during recycling of ∼6800 J. The cryostat can operate without any relevant changes to performance tilted down to 50° from the vertical position, as needed at the telescope focal plane.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.