Abstract
A two-compartment gel phantom for VOI profile measurements in volume-selective spectroscopy in small-bore units is presented. The phantom is cylindrical with two compartments divided by a very thin polyethene film. This thin film permits measurements with a minimum of susceptibility influences from the partition wall. The phantom was used for evaluation of the volume selection method ISIS (image-selected in vivo spectroscopy). The position of the phantom was fixed in the magnet during the measurements, while the volume of interest (VOI) was moved stepwise over the border. The signal from the two compartments was measured for each position and the data were evaluated following differentiation. We have found this phantom suitable for VOI profile measurements of ISIS in small-bore systems. The phantom forms a useful complement to recommended phantoms for small bore-spectroscopy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.