Abstract

The GE FASTlab radiosynthesis module is routinely used for the production of [(18)F]FDG, utilizing the commercially available phosphate cassettes. Because of the observation of a white precipitate in the product vial before the product expiry time, we re-examined the quality of the produced [(18)F]FDG solution. Phosphate buffered [(18)F]FDG solution was synthesized on the FASTlab and analyzed at both National Taiwan University Hospital (NTUH) of Taiwan and Royal Brisbane and Women's Hospital (RBWH) of Australia. In addition to the standard product quality control (QC), the concentration of aluminum (Al(3+)) as probable cause of the precipitations in the [(18)F]FDG solution was analyzed by inductively coupled plasma mass spectrometry (ICP-MS at RBWH) and inductively coupled plasma optical emission spectrometry (ICP-OES at NTUH), and using three semi-quantitative methods at NTUH, Advantec(®) Alumi Check Test Strip, Quantofix(®) Aluminum Test Strip and MColortest™ Aluminum Test kit. The precipitates were observed in the [(18)F]FDG solution within 24 (NTUH) and 6 (RBWH) hours after the end of synthesis in 38-100% of the batches, dependent on the batch of the FASTlab cassettes. Addition of metal-free HCl(aq) to aliquots of [(18)F]FDG containing precipitate, followed by ICP-MS analysis revealed Al(3+) concentrations of 70-80ppm. Al(3+) concentrations of 10-12ppm were detected in [(18)F]FDG batches that did not show any precipitation. In contrast, less than 5ppm of the residual Al(3+) was detected by semi-quantitative methods in all batches. The US (USP), British (BP), European (EP) and Japanese (JP) pharmacopeias demand that [(18)F]FDG for injection should be clear and particulate free within the given shelf-life/expiration time. To avoid Al-phosphate precipitation within the product expiry time, FASTlab citrate cassettes, rather than phosphate cassettes, should be used for [(18)F]FDG production. Although testing for Al(3+) is not listed in the [(18)F]FDG monographs of the USP, BP and EP, residual Al(3+) levels should be considered in the interests of patient safety.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.