Abstract

The sensitivity of a Stark-modulated waveguide microwave spectrometer can be significantly improved by using a travelling wave tube or solid state microwave amplifier prior to the crystal detection. In this paper the authors discuss the travelling wave tube. The increase in the sensitivity results from the improved noise figure of the travelling wave tube relative to crystal video type detectors. Wide bandwidth travelling wave tube or solid state amplifiers are available so full waveguide bandwidths may be utilised. Rotational transitions for J=1 from 0 lines of cyanogen bromide, BrCN, were measured with the travelling wave tube system and a standard video detector and the spectra are given here.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.