Abstract

Standard gas mixtures (SGMs) of trace level in high-pressure cylinder are typically prepared by multi-stage dilution from the pure gases. Here ‘a small amount of gas transfer technique’ was developed to prepare SGMs, in which the number of dilution stages could be much less than typical process. The preparation process and the technique were introduced in detail. And the theory and key points of the technique were discussed. As an example, Freon 12 SGMs of trace level (100nmol/mol) were prepared by the technique through only three-stage dilution, the relative standard uncertainty of preparation was less than 0.2%. The results showed consistency agreement of <0.2%. In order to verify the reliability of the technique a Freon 12 SGMs of 1μmol/mol and 0.1μmol/mol were prepared by conventional method through four-stage and five-stage dilution respectively. The standard uncertainty of Freon 12 prepared by two methods was similar and the result of the consistency of two methods was satisfied. Considering the savings of time and labor, the technique is worth to be recommended.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.