Abstract
To describe a new device for the transcervical collection of amniotic fluid (AF) in patients with ruptured membranes, and to compare the concentration of proteins in fluid retrieved by transabdominal amniocentesis and the transcervical AF collector. Paired AF samples were collected in patients with preterm prelabor rupture of membranes (PROM) (n=11) by transabdominal amniocentesis and with the transcervical AF collector (Yoon's AF Collector™). Three proteins known to have high concentrations in AF [α-fetoprotein (AFP), β-human chorionic gonadotrophin (β-hCG), and prolactin] were measured. (1) There was a significant correlation between the concentrations of analytes in AF obtained by transabdominal amniocentesis and by the transcervical AF collector (r=0.94, P<0.001 for AFP; r=0.96, P<0.001 for β-hCG; r=0.72, P<0.05 for prolactin); (2) Bland-Altman plots showed no evidence of heteroscedasticity between transabdominal or transcervical AF concentrations of these markers. There was a strong correlation between the concentrations of proteins in AF collected by amniocentesis or with the transcervical device.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.