Abstract

A time-of-flight positron emission tomography (TOF-PET) detector was developed based on a 6×9 LYSO array and four single channel photomultipliers (Hamamatsu R9800). Leading-edge timing circuit with optimized parameter was used instead of the constant fraction discriminator. The results showed that all 54 elements in the flood histogram could be identified clearly. The average coincidence resolving time was 402 ps FWHM.