Abstract
Liver disease is a great danger to human health. The determination of blood level of Cholyglycine acid (CG) is a vital biomarker for the assessment of liver function in clinic, which is contribute to the diagnosis of liver diseases. Thus, establishing accuracy, rapid and convenient method for the detection of glycolic acid is of great significance. In this study, a time-resolved-fluorescence (TRF) lateral flow assay for rapid detection of CG was development. The analytical detection limit (mean of zero-2 SD) was 0.06 μg/mL The method showed good linearity in the range of 0.2–40 g/mL and was not affected by biomolecules with similar structure to CG. The analytical mean recovery of control was between 90–110% and the imprecision of intra- and inter-assay of CVs was less than 10%. No significant matrix effect was observed in saline, serum, plasma or whole blood. A good correlation was found with the homogeneous enzyme immunoassay (HEIA) assay (slopes 1.0463, y-intercepts 0.2721 μg/mL, R = 0.989, n = 50, P < 0.001). The CG TRF analysis could provide reproducible and quantitative information about the state of liver in a few minutes, which is suitable for the detection of liver diseases in point-of-care-testing (POCT) conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.