Abstract

We have used Doppler echocardiography to estimate the stroke volume (SV) in a study of 13 patients equipped with DDD pacemakers. SV was measured both during DDD and VVI pacing after observation times of 1,3,6, and 12 months of DDD pacing. SV was also measured at seven atrioventricular (AV) intervals (75-250 ms) in the search for optimal AV intervals. Mitral flow velocity was investigated to see if DDD pacing resulted in synchronous atrial contraction, and if mitral insufficiency existed at any of the pacing modes. Compared with the VVI mode, DDD pacing resulted in a mean increase in SV of 21 +/- 2% for the four observation periods. Two patients with severe left ventricular failure had no significant increase in SV during DDD vs VVI pacing. In each patient, an optimal AV interval ranging between 100-250 ms for the SV was found. Velocity profiles of mitral flow showed synchronous atrial contraction during DDD pacing, but not during VVI pacing. Mitral insufficiency was not seen in any pacing mode. DDD pacing resulted in a reduction in SV during the first 6 months, and was constant thereafter. Doppler echocardiography can be used repeatedly to evaluate the hemodynamic response of DDD pacing vs VVI pacing, and to find which AV interval gives the highest SV in the individual patient. Our study further shows that the hemodynamic benefit of DDD pacing is present after short-term as well as after long-term DDD pacing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.