Abstract

Capillary electrophoresis DNA sequencing has been accomplished by using four different energy transfer primers and three fluorescence detection channels. Methods have also been developed to deconvolve the three-color data into the four base concentrations. The nonnegative least squares and model selection method resulted in the best accuracy. The three-color data were compared to sequencing data obtained using four detection channels and four energy transfer primers. The average accuracy rates obtained over three 500 base M13mp18 runs using three-color coding were 96% including 18 uncallable compressions and 99.6% if these compressions are excluded. The average accuracy rate obtained using four-color coding was 96.3% including 18 uncallable compressions and 99.9% if these compressions are excluded. Although it is unlikely that three-color schemes will replace four-color sequencing, these methods have exposed basic concepts that will be useful in the development of higher-order multiplex coding methods for DNA analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.