Abstract
In order to create a less thrombogenic heart valve with improved longevity, a prosthetic heart valve was developed using thin film nitinol (NiTi). A “butterfly” thin film NiTi valve was constructed using a single, elliptical piece of thin film NiTi and a scaffold made from Teflon tubing and NiTi wire. Flow tests and pressure readings across the valve were performed in vitro in a pulsatile flow loop. Biocorrosion experiments were conducted on untreated and passivated thin film nitinol. To determine the material’s in vivo biocompatibility, thin film nitinol was implanted in a pig using a stent covered with thin film NiTi. Flow rates and pressure tracings across the valve were comparable to those through a commercially available 19 mm Perimount Edwards tissue valve. No signs of corrosion were present on samples of thin film nitinol after immersion in Hank’s solution for 1 month. Finally, organs and tissue samples explanted from the pig 17 days after thin film NiTi implantation appeared without disease, and the thin film nitinol itself was without thrombus formation or endothelialization. Although long term testing will be needed, thin film NiTi may be very well suited for use in artificial heart valves.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.