Abstract

Medical X-ray digital imaging systems such as mammography, radiography and computed tomography (CT), are composed from efficient radiation detectors, which can transform the X-rays to electron signal. Scintillators are materials that emit light when excited by X-rays and incorporated in X-ray medical imaging detectors. Columnar scintillator, like CsI:T1 is very often used for X-ray detection due to its higher performance. The columnar form limits the lateral spread of the optical photons to the scintillator output, thus it demonstrates superior spatial resolution compared to granular scintillators. The aim of this work is to provide an analytical model for calculating the MTF, the DQE and the emission efficiency of a columnar scintillator. The model parameters were validated against published Monte Carlo data. The model was able to predict the overall performance of CsI:Tl scintillators and suggested an optimum thickness of 300μm for radiography applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.