Abstract
A complete, scalar diffraction based, mathematical analysis Is presented for evaluating the response of both ordinary (reflectance) and magneto-optic confocal and non-confocal scanning laser microscopes. Magneto-optic configurations employing both single and differential detector arrangements are analyzed. A transfer function approach is adopted that enables the imaging properties of the optical system to be separated out from the properties of the object. It is therefore straightforward to compare quantitatively the imaging performance of different optical arrangements. A particularly interesting result is that the impulse response of a magneto-optic scanning laser microscope with the commonly used differential detection technique is identical to that of a simple incoherent reflectance microscope.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.