Abstract
The aim of this study was to examine the angular distribution of the light emitted from radiation-excited scintillators in medical imaging detectors. This distribution diverges from Lambert's cosine law and affects the light emission efficiency of scintillators, hence it also affects the dose burden to the patient. In the present study, the angular distribution was theoretically modeled and was used to fit experimental data on various scintillator materials. Results of calculations revealed that the angular distribution is more directional than that predicted by Lambert's law. Divergence from this law is more pronounced for high values of light attenuation coefficient and thick scintillator layers (screens). This type of divergence reduces light emission efficiency and hence it increases the incident X-ray flux required for a given level of image brightness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.