Abstract
Bipolar electrodes are potentially useful for a variety of sensing applications, but their implementation has been hampered by an inability to easily monitor the current through such electrodes. However, current can be indirectly determined using electrogenerated chemiluminescence (ECL) as a reporting mechanism. This paper provides a detailed theoretical analysis of ECL reporting at bipolar electrodes. In addition, experiments are described that confirm the theory. Finally, we correlate ECL intensity directly to current through the use of split bipolar electrodes. The results indicate that the lowest current that can be indirectly detected through ECL reporting is ∼32 μA/cm2, which corresponds to a reporting sensitivity of ∼7200 counts/nA in the present experimental system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.