Abstract

Calibration and use of the diffusion porometer are imprecise because of imperfect understanding of steady diffusion through a porous material. The case of a flat plate with uniformly distributed right circular cylindrical holes is approximated by diffusion through two finite right circular cylinders: one representing the pore, and one representing the area into which vapor diffuses — its size determined by the mutual interference with neighboring pores. An exact three-dimensional solution is presented. It is found that for large pore spacing the empirical result of Holcomb and Cooke is excellent, but for close pore spacing some error occurs. A method for calculation of true calibration plate resistance is described as well as a method for estimating pore size for an unknown plate or membrane.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.