Abstract

High resolution light microscopical images that can be superimposed on a test system for the estimation of stereological parameters are obtained by a television system modified for point-counting morphometry. The full resolving power of the light microscope is exploited and a test system applied in the same plane as the microscopical image displayed on the monitor screen. Volume fraction estimates of nuclei and zymogen granules from the rat pancreas made with the television microscopical system and with electron-microscopic micrographs rendered the same precision, and showed that the volume fraction of objects as small as zymogen granules was measurable with the light microscope. The measuring with the television microscopical system was about 15 times faster than with electron-microscopic micrography and about 5 times faster than with conventional photomicrography and the light microscope.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.