Abstract

In this article, we present a microfluidic technique for the rapid enumeration of bacterial density with a syringe filter to trap bacteria and the quantification of the bacterial density through pressure difference measurement across the membrane. First, we established the baseline differential pressure and hydraulic resistance for a filtration membrane by fully wetting the filter with DI water. Subsequently, when bacteria were infused and trapped at the pores of the membrane, the differential pressure and hydraulic resistance also increased. We characterized the infusion time required for the bacterial sample to achieve a normalized hydraulic resistance of 1.5. An equivalent electric-circuit model and calibration data sets from parametric studies were used to determine the general form of a calibration curve for the prediction of the bacterial density of a bacterial sample. As a proof of concept, we demonstrated through blind tests with Escherichia coli that the device is capable of determining the bacterial density of a sample ranging from 7.3 × 106 to 2.2 × 108 CFU/mL with mean and median accuracies of 87.21% and 91.33%, respectively. The sample-to-result time is 19 min for a sample with lower detection threshold, while for higher-bacterial-density samples the measurement time is further shortened to merely 8 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.