Abstract

In the first place the principle and properties of a hood method for measurement of metabolism during sleep are described. In the next place the hood system built by authors as a trial is described. This system consists of a hood, an electric blower, a wet gas meter and 300-liter gas bags. The dead space of the hood is about 10 liters. The blow rate was adjusted at 25, 30 and 40 l/min. The VO2 and VCO2 values were measured for 3 adult men, the correlation coefficients between the hood method and the Dougles bag method were 0.978 and 0.980, respectively. No relationship appeared between blow rate and accuracy of measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.